Skip to content

Innovative disposable pipette extraction for concurrent analysis of fourteen psychoactive substances in drug users' sweat.

Nayna Cândida Gomes, Oscar G Cabrices, Bruno Spinosa De Martinis

Journal of chromatography. A August 16, 2024 DOI: 10.1016/j.chroma.2024.465136 via PubMed

Summary

AI-generated from the abstract

A new method using disposable DPX-SCX extraction tips and gas chromatography-mass spectrometry can simultaneously detect and quantify 16 psychoactive substances—including amphetamines, cocaine, fentanyl, and several novel psychoactive substances—in sweat samples collected with laboratory-developed patches. Validation showed quantification limits between 2 and 30 ng per patch and linearity from 2 to 1100 ng per patch. When applied to 30 sweat samples from volunteer drug users, the method detected cocaine, MDMA, nicotine, caffeine, and other compounds, with recovery rates of 72.4% to 97.1%. Sweat is a viable, non-invasive biological matrix for monitoring drug use, particularly in rehabilitation settings.

Study at a glance

Characteristics Method validation and application Peer reviewed
Sample size 30
Population Volunteer drug users
Keywords Dpx-scx Gc-ms Psychoactive substances Sweat Drug_testing substance screening
Citations 12
Key finding The validated DPX-SCX extraction and GC-MS method effectively quantifies multiple psychoactive substances in sweat, with recovery rates from 72.4% to 97.1% and limits of quantification from 2 to 30 ng per patch.

Abstract

Widespread consumption of drugs of abuse worldwide has caused concern: it adversely affects public health, individual safety, and social structures. Experts are particularly alarmed because new psychoactive substances have been increasingly detected in biological samples. In recent years, several studies have focused on developing methods to identify psychoactive substances in alternative biological matrices, such as sweat. This approach holds promise for monitoring substance use, especially in individuals undergoing rehabilitation. Among the commonly employed analytical procedures, extraction using disposable DPX tips stands out as a novel, miniaturized, and promising technique. This study aimed to validate and to apply a method to analyze various substances, including amphetamine, methamphetamine, MDMA, MDA, MDEA, cocaine, cocaethylene, anhydroecgonine methyl ester, dibutylone, N-ethylpentylone, 25E-NBOMe, 25CNBOMe, 2CC, 2C-E, fentanyl, and carfentanil, in sweat samples simultaneously. In this method, sweat is collected by using laboratory-developed patches, and extraction is conducted with DPX-SCX tips. Gas chromatography coupled to mass spectrometry is employed to separate, to identify, and to quantify the analytes. Validation results indicated that the quantification limit ranged from 2 to 30 ng of analyte/patch, and that the method was linear for analyte concentrations ranging from 2 to 1100 ng/patch. The validated method was applied to analyze 30 sweat samples collected from volunteers drug users and processed by using both the selected ion mode (SIM) and full scan. The method was able to detect and to quantify substances such as cocaine, cocaethylene, anhydroecgonine methyl ester, MDMA, MDA, nicotine, cotinine, caffeine, procaine, lidocaine, and ethylamphetamine simultaneously. The recovery rates ranged from 72.4 % to 97.1 %. The analytes were stable in the biological matrix. In conclusion, the validated method proved effective and allowed the target analytes to be quantified in sweat samples, highlighting that sweat is a viable matrix for analyzing drugs of abuse.

Comments

No comments yet.

Log in to comment